Protocols - Wohlwend
Start up
- Open nitrogen gas valve
- Press yellow "System Dry Up" button. Run gas through for 30 minutes.
- Press green "On/Off" button
- Press green "Nitrogen" button
- Open liquid nitrogen tank
- Wait until red "Nitrogen Level" light goes out
- Press yellow "Drive in" button.
- Yellow light will start flashing
- "Piston Down" red lamp will go on
- Oil pressure will increase to 100 bar
- "Drive In" light will go out
- Press green "Auto" button.
Use
- Insert holder and safety pin
- Green "Jet Ready" lamp should light
- Press yellow "Jet Auto" button
- Wait for yellow "Jet Over" lamp
- Pull pin and remove sample
Measuring temperature and pressure (traces to check performance)
*
Shut down
- Close liquid nitrogen tank
- Press green "On/Off" Button
- Press yellow "System Dry Up" button * Light will blink for 10 hours while system drys
- Once lamp stops flashing close nitrogen gas valve
-- KdDerr - 03 Oct 2007
|