Protocols - Edwards 306 Vacuum Evaporator
Important:
- Please report any bell jar damage. Failure to do so may risk serious injury to others.
General Notes
- Wear gloves when touching sources
- Positive source-ceramic
- Negative source- no ceramic
- It is critical that pos+ is connected to pos+ and neg- to neg- when attaching accessories.
Start Up
- Switch instrument power on (above Rotatilt knob) turn knob from 0 to 1
- Error message will be displayed press reset then start
- Wait for sealed light to appear
Venting the chamber
- Press vent
- Unhinge bell jar
- Carefully lift bell jar
- Add or remove samples and/or accessories
Pumping down the chamber
- Use compressed air to clean pirani before pumping down
- Carefully close the bell jar
- Lock the safety hinge
- Press cycle to begin pump down
Shut down
- Close bell jar
- Press Cycle and wait until high vacuum has been reached
- Press seal
- Press stop
- When the digital display reads STANDBY turn power switch to 0
Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM
-- DavidStokes - 18 Mar 2005
|